DT600 - CE05 Troubleshooting

If you're encountering CE05 error as soon as you turn on the machine or during the count, please proceed with the following steps

Step 1: Please clean the Start Sensor located on the Hopper Stage.

Step 2: Check the Start Sensor voltage

1. Long press the ADD button to access the Voltage Dashboard
2. Cover the Start Sensor with your finger

If the voltage goes up >300 and back to the 0-5 range when you remove your finger the Start Sensor is good.

If the voltage does not move, please contact support@deteckusa.com for further assistance

Step 3: Check if the start sensor cables are properly connected
Article: https://deteckusa.reamaze.com/articles/dt600-start-sensor-cables-connection